Products Search
>
HPLC
>
System Controller
>
Error Code:[0x2F03]
Back
No : 3769
Date of Issue : 2020/12/02 16:43
Print
Error Code:[0x2F03]
Error Message: System controller failed to communicate with the instrument. After checking the communication environment, turn off the main power and restart.
Category :
Products Search
>
HPLC
>
System Controller
Answer
«Main Cause»
A communication error occurred with instrument.
«Countermeasure»
Check the optical link cable, turn off the main power of the system controller and each instrument, and restart it.