«Main Cause»
Temperature control is turned off because the temperature of the column oven exceeded the maximum temperature.
«Countermeasure»
Settings the instrument parameter so that the difference between the set value of the oven temperature and [upper temperature limit] is about 5ºC. If the difference is small, the difference may exceed [upper temperature limit] before the temperature of the column oven becomes set value.