• No : 3683
  • Date of Issue : 2020/12/02 16:31
  • Print

Error Code:[0x8F01]

Error Message: An empty test tube is missing. Check the end tube number Settings.
Category : 

Answer

«Main Cause»
No empty vial in the fraction collector.
 
«Countermeasure»
(1) Fraction collector has used up to the final vial. Check the final vial (F.VIAL) setting.
(2) Follow the procedure below.
(a) Set empty vials on the rack.
(b) Create a dummy method (of which I.VIAL="0", analysis time = 0.01min) and run the dummy analysis. Then the memorized vial position is cleared, and the fraction collector will use the vial No.0.
(3) Run the batch from the interrupted line.