Error Code:[0x8967]
Error Message: The wavelength of the installed cell is not calibrated. Perform wavelength accuracy calibration.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
The wavelength of the installed cell is not calibrated.
«Countermeasure»
Perform wavelength calibration.