• No : 3559
  • Date of Issue : 2020/12/02 16:30
  • Print

Error Code:[0x0012] ERR NDLE HOME Z

Error Message: Needle Z direction home position failed to detection.
Category : 

Answer

«Main Cause»
Move to Z (up and down) direction was faulty.
 
«Countermeasure»
Check if nothing is stuck in the movable part.
If nothing is stuck, contact a Shimadzu sales or service representative.