Error Code:[0x1003]
Error Message: The suppressor-related parameter Settings is incorrect. Check on Settings.
Category :
Products Search
>
HPLC
>
System Controller
Answer
«Main Cause»
The suppressor-related parameters are not correct.
«Countermeasure»
Check the values and set so that values are correct.