Error Code:[0x1002]
Error Message: Since the suppressor is running, operation is not possible.
Category :
Products Search
>
HPLC
>
System Controller
Answer
«Main Cause»
Full-regeneration was operated while HPLC system was activated.
«Countermeasure»
Do not operate full-regeneration while HPLC system is activated.