Products Search
>
HPLC
>
System Controller
>
Error Code:[0x9004]
Back
No : 3497
Date of Issue : 2020/12/02 16:10
Print
Error Code:[0x9004]
Error Message: The backup parameters of the controller are corrupted. The method file and sequence file were initialized.
Category :
Products Search
>
HPLC
>
System Controller
Answer
«Main Cause»
The method and sequence are returned to the same condition as when the instrument was last started because the backup parameters are corrupted.
«Countermeasure»
Check the method and sequence.