(EHF) Control switching is not possible

(EHF) Control switching is not possible
EHF-E, U, L Series
Servo Controller 4830
Category : 

Answer

If you cannot solve this problem by checking possible causes and taking appropriate actions described below, contact a Shimadzu sales or service representative.
 
Cause1: The sensor cable is not connected, or almost disconnected.
Action1: Connect the sensor cable securely.
 
Cause2: The load cell cable is disconnected, or almost disconnected.
Action2: Connect the cable securely. A warning message is displayed when the servo controller starts.
 
Cause3: Select “Invalid”.
Action3: Please select "Valid".
  Click here for "(EHF) TD calibration".