(EHF) Out of control

(EHF) Out of control
EHF-E, U, L Series
Servo Controller 4830
Category : 

Answer

 
Cause1: The sensor cable is not connected, or almost disconnected.
Action1: Connect the sensor cable securely.
 
Cause2: The elongation meter is improperly mounted.
Action2: Mount the elongation meter securely.
 
Cause3: The sensor polarity is reserve.
Action3: Correct the sensor polarity.
 
Cause4: The servo valve cable is disconnected, or almost disconnected.
Action4: Connect the servo valve cable securely.
 
Cause5: LOAD operation is not executed. (For hydraulic system)
Action5: Press the [LOAD] button to execute LOAD operation.
 
Cause6: Hydraulic pressure is not increased. (For hydraulic system)
Action6: Increase the hydraulic pressure to an appropriate value.
 
Cause7: Piston is in the cushion portion outside the rated range . (When machine did not work for a long period, piston may does not move in a jog dial.)
Action7: Raise Hydraulic slowly to 10MPa (rated pressure in case of the low pressure type ). Then return piston into rated range using "Mean value moving".