CARm calc. ratio out of range[Error Code:5075 5076 5108 5109]
The carrier gas total flow calculated from split ratio program, etc. is outside the set range.
- Category :
Answer
Change the split ratio so that the following equation is satisfied.
(Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow)
(Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow)