• Font Size
  • S
  • M
  • L
  • No : 5916
  • Date of Issue : 2021/05/18 15:37
  • Last Updated : 2021/06/12 01:21
  • Print

DetAPCn hydrogen flowing over the specified value [Error Code:2300 2301 2302 2303]

Hydrogen flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation.
Category : 

Answer

  • Do not open or close the main valve of the supply gas during APC operation.
  • Restart the GC with the gas supplied.
  • Check that the hydrogen gas supply pressure is within the range below.
        
  • Check that the gas supply pressure is stable.
  • If the error appears repeatedly even after checking the above, the control system of the APC may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.