• Font Size
  • S
  • M
  • L
  • No : 4712
  • Date of Issue : 2020/12/17 16:43
  • Print

(GCMS) "Detector saturated" message appears during analysis

(GCMS) "Detector saturated" message appears during analysis.
Category : 

Answer

1. Turn on the filament after solvent has completely eluted. If peaks elute before the solvent, use a time program to turn the filament off and on.
2. Decrease the detector voltage.
3. Reduce the sample concentration by making a dilution or reduce the injection volume.
4. If unavoidable, switch the filament OFF.

May I have your opinion?

Please provide your feedback below. If a reply is required, please use one of the four red inquiry buttons at the bottom of the screen.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.