- Products Search > HPLC > System > Ion Chromatograph > Noises or fluctuation became large (Ion Chromatograph)
Noises or fluctuation became large (Ion Chromatograph)
Noises or fluctuation became large (Ion Chromatograph)
Noise: 4 to 7 nS/cm
- Category :
Answer
Bubbles have accumulated. Pass water through the suppressor.
When a large noise persists even after taking the action above, perform "Regeneration Procedures".
When a large noise persists even after taking the action above, perform "Regeneration Procedures".
*The small fluctuations are due to the continuous electrodialysis but not due to a problem with the suppressor.
Click here for Regeneration Procedures.
Click here for Baseline of the Electrodialytic Suppressor.