Optional products connected to the instrument are not displayed
Optional products connected to the instrument are not displayed.
Example:
· The optional value connected to the pump is not displayed.
· The flow channel selection valve for the column oven is not displayed.
· The solvent recycle valve connected to the detector is not displayed.
· The sub-controller valve is not displayed.
· The optional value connected to the pump is not displayed.
· The flow channel selection valve for the column oven is not displayed.
· The solvent recycle valve connected to the detector is not displayed.
· The sub-controller valve is not displayed.
- Category :
Answer
«Main Cause»
The [Used]/[Not Used] settings for various optional products in configuration settings are set to [Not Used].
«Countermeasure»
Make sure configuration settings for each instrument are specified correctly.