Instrument is not recognized when connected via an optical link cable to an SCL-40 unit not connected to workstation software
Instrument is not recognized when connected via an optical link cable to an SCL-40 unit not connected to workstation software.
- Category :
Answer
«Main Cause»
Link mode setting is [Fixed]
(tends to occur if workstation software was force-quit).
(tends to occur if workstation software was force-quit).
«Countermeasure»
Set link mode to [Auto].