- Products Search > HPLC > Column Oven > Error Code:[0x940E] WARNING LOW SET TEMP
Error Code:[0x940E] WARNING LOW SET TEMP
Error Message: Column oven: The temperature set currently is too low compared to the room temperature. Change setting temperature because it may not reach set temperature.
- Category :
Answer
«Main Cause»
The setting is too low, relative to room temperature.
«Countermeasure»
Check the column oven temperature setting in instrument parameter settings. If it is more than 12ºC from the room temperature, change the setting so that it is within 12ºC from the room temperature.