Error Code:[0x940B] ERROR OVER T.MAX

Error Message: Column oven: The temperature control was stopped because the temperature exceeded upper temperature limit. Check the setting value of Maximum Temperature [T.MAX] on the column oven.
Category : 

Answer

«Main Cause»
Temperature control was stopped because the column oven temperature exceeded the upper limit temperature.
 
«Countermeasure»
Set the column oven temperature setting in instrument parameter settings to about 5ºC from the [T.MAX] setting value. A small difference in the settings could cause the column oven temperature to exceed the [T.MAX] value before it reaches the setting value.