Error Code:[0x5404] ERROR HIGH SET TEMP

Error Message: Column oven: The temperature set currently is too high. The high-temperature of flow-line selection valve or Column Management Device (CMD) cannot be higher than 85ºC during connecting. Change the temperature to 85ºC or lower.
Category : 

Answer

«Main Cause»
The temperature setting is too high. The temperature cannot be set higher than 85ºC if a flow channel selection valve or column management device (CMD) is connected.
 
«Countermeasure»
Set the temperature to 85ºC or less or disconnect the CMD.