FAQ - Frequently Asked Questions
- Products Search > HPLC > Column Oven > Error Code:[0x5400] ERROR INT-VALVE1 [0x##]
Error Code:[0x5400] ERROR INT-VALVE1 [0x##]
Error Message: Column oven: The built-in flow-line selection valve 1 is restarted because an error is detected in the valve.
- Category :
Answer
«Main Cause»
An error was detected in the flow channel selection valve 1 inside the column oven.
(The error code "0x##" indicates a valve unit error. Refer to the valve unit instruction manual.)
«Countermeasure»
When [CE] is pressed, the valve unit restarts automatically.
If an occurs again after the valve unit restarts, switch the instrument power OFF and contact a Shimadzu sales or service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.