  • No : 3852
  • Date of Issue : 2020/12/02 16:43
  • Print

Error Code:[0x5F02] WARNING NO PAIRING

Error Message: When the operation panel was replaced, initial settings were not set correctly. The backup function is not working properly.
Category : 

Answer

«Main Cause»
Default settings were not set after operation panel replacement.
 
«Countermeasure»
The system can continue to be operated as it is, but the instrument information backup functionality is disabled, so that functionality for restoring lost data cannot be used. Contact a Shimadzu sales or service representative.