Error Code:[0x2F35]

Error Message: System controller failed to communicate with instrument. After checking the communication environment, turn off the main power and restart. Check the communication environment.
Category : 

Answer

«Main Cause»
A communication error occurred with the instrument.
 
«Countermeasure»
Check the optical link cable.
If that does resolve the error, switch OFF the main power for both the system controller and applicable instrument and restart them.