- Products Search > HPLC > Column Oven > Error Code:[0x0070] ERR OVER T.MAX
Error Code:[0x0070] ERR OVER T.MAX
Error Message: Column oven: The temperature control was stopped because the temperature exceeded the upper temperature limit. Check the setting value of Maximum Temperature(T.MAX) on the column oven.
- Category :
Answer
«Main Cause»
Temperature control is turned off because the temperature of the column oven exceeded the maximum temperature.
«Countermeasure»
Settings the instrument parameter so that the difference between the set value of the oven temperature and [upper temperature limit] is about 5ºC. If the difference is small, the difference may exceed [upper temperature limit] before the temperature of the column oven becomes set value.