- Products Search > HPLC > Column Oven > Error Code:[0x0077] WARNING: LOW SET TEMP
Error Code:[0x0077] WARNING: LOW SET TEMP
Error Message: The set value of oven :oven temperature is too low compared to room temperature.
- Category :
Answer
«Main Cause»
For CTO-10AC (vp) or CTO-20AC: If the COOLER MODE is "1" or "2", the room temperature is less than -10ºC. When "0" is set, Settings of oven temperature below room temperature.
For CTO-10A(vp): Settings of oven temperature to room temperature or less.
CTO-10 ASvp: Settings of oven temperature below -15ºC.
«Countermeasure»
Temperature control does not stop, but the set temperature is out of specification-warranty range. Raise set temperature to the appropriate temperature. If you do not change the set temperature, the warning message is display periodically.