FAQ - Frequently Asked Questions
- Products Search > HPLC > System Controller > Error Code:[0x9071]
Error Code:[0x9071]
Error Message: The external detector set on the instrument is not connected.
- Category :
Answer
«Main Cause»
Acquisition was started when a detector that differs from the detector set in the simple parameters of the acquisition main screen was connected.
«Countermeasure»
Either connect the detector that is set in the simple parameters of the acquisition main screen, or change the settings in the simple parameters of the acquisition main screen to match the connected detector.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.