Error Code:[0x5630] WARNING LOW SET TEMP
Error Message: Chemical reactor: The temperature set currently is too low compared to the room temperature. Change setting temperature because it may not reach set temperature.
- Category :
Answer
«Main Cause»
The setting is too low, relative to room temperature.
«Countermeasure»
Check the oven temperature setting in instrument parameter settings. If the setting is lower than room temperature +10ºC, change the setting to be higher than room temperature +10ºC.