Error Code:[0x562D] ERROR GAS DETECTED
Error Message: Chemical reactor: The gas sensor detected organic solvent gas, so the temperature control was stopped. Check leakage of fluid from the flow line, such as the tubing connection.
- Category :
Answer
«Main Cause»
Temperature control was stopped because the gas sensor inside the oven detected organic solvent leakage.
«Countermeasure»
(1) If an organic solvent leaks inside the oven, implement corrective measures to prevent tubing leaks.
(2) Check for organic solvent gases around the instrument. If there are any organic solvent gases, thoroughly ventilate the room.
(3) If there are no problems with the above two measures, then calibrate the gas sensor sensitivity.