FAQ - Frequently Asked Questions
- Products Search > Columns, Reagents and Consumables > LC Columns > (LC Column) The back pressure of the column is higher tha...
(LC Column) The back pressure of the column is higher than when it was used previously.
The back pressure of the column is higher than when it was used previously.
- Category :
Answer
->Make sure that the column is stored according to the instructions in the manual.
->Please check that the mobile phase containing salt such as buffer solution is not flowing immediately after connecting to the equipment.
->Salt may precipitate and the column may be clogged. If precipitation of salt is suspected, slowly pass a solvent with high elution power such as pure water and try to improve. If the problem is not resolved, please replace the column to new one.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.