FAQ - Frequently Asked Questions
- Products Search > HPLC > LC Columns > (LC Column) Poor reproducibility of retention time betwee...
(LC Column) Poor reproducibility of retention time between days
Retention time are shifted between days even if using same analytical conditions (column, mobile phase, sample, system, and etc.).
- Category :
Answer
->The composition may have changed due to volatilization of the mobile phase or sample. Reprepare and check again.
->Ensure that the column is aged sufficiently.
->Depending on the storage method, the column may have deteriorated. In this case, replace the column.
->For the recommended column storage method, see the instruction manual.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.