FAQ within [ Ion Chromatograph ]
1-7 of Total 7
-
Compared to the chemisorption-based cartridge suppressor,...
«Cause» Peak detection parameters are incorrect. «Countermeasure» The baseline is more unstable with this instrument than the chemical cartridge... Show Detail
- No:4382
- Date of Issue:2020/10/26 13:26
-
Leakage from suppressor (Ion Chromatograph)
When the suppressor is left unused, the inner side dries to cause leakage from the filter and the flow line. In this case, perform "Hydration Procedure". Whe... Show Detail
- No:4383
- Date of Issue:2020/10/26 13:28
-
Baseline of the Electrodialytic Suppressor (Ion Chromatog...
The electrodialytic suppressor is characterized by the baseline varying every moment. The baseline variation is numerically represented by a noise value. The... Show Detail
- No:4391
- Date of Issue:2020/10/26 13:48
-
Hydration Procedure of ICDS (Ion Chromatograph)
The suppressor is ready for proper operation when the ion exchange membranes and mesh absorb water and swell (hydrated). Hydrate the suppressor in the fol... Show Detail
- No:4388
- Date of Issue:2020/10/26 14:03
-
Regeneration Procedure of ICDS (Ion Chromatograph)
Continuing passing liquid through the suppressor without synchronizing the suppressor current with the pump causes the ion exchange membranes of the suppress... Show Detail
- No:4389
- Date of Issue:2020/10/26 14:23
-
Drift does not stabilize (Ion Chromatograph)
Hydrate the suppressor and use it as usual. Delivering eluent without hydrating the suppressor may cause sodium ion to enter a section from which the ion is... Show Detail
- No:4387
- Date of Issue:2020/10/26 14:24
-
Noises or fluctuation became large (Ion Chromatograph)
Bubbles have accumulated. Pass water through the suppressor. When a large noise persists even after taking the action above, perform "Regeneration Procedure... Show Detail
- No:4386
- Date of Issue:2020/10/26 14:27