FAQ - Frequently Asked Questions
- Products Search > Non-Destructive Testing (NDT) Systems
Keyword Search
Non-Destructive Testing (NDT) Systems
Frequently Referenced FAQ
FAQ within [ Non-Destructive Testing (NDT) Systems ]
-
(SMX-225CT) Rotating the X-ray Tube Head Tungsten Target
This system emits X-rays by irradiating a target with a focused electron beam. The tungsten can melt and vaporize, caused a rough targetsurface, which can... Show Detail
- No:8680
- Date of Issue:2022/05/23 09:11
- Category: inspeXio SMX-225CT
-
(SMX-225CT) Refilling Cooling Water
1. Open the lid of the cooling water unit. 2. Turn the lid on the top of the cooling water tank counterclockwise to open it. 3. Check that... Show Detail
- No:8681
- Date of Issue:2022/05/23 09:14
- Category: inspeXio SMX-225CT
-
(SMX-225CT) Create Calibration Table
1. The [Password] window is displayed when the [Management] button in CTsoftware inspeXio64 is clicked. 2. Enter the system manager password ("00... Show Detail
- No:8682
- Date of Issue:2022/05/23 11:26
- Category: inspeXio SMX-225CT
- Brached FAQ
-
(SMX-225CT) Error Detection on Flat Panel Detector
If any abnormality occurs on a device that captures image data, such as the flat paneldetector and the command processor which retrieves image data from the ... Show Detail
- No:8689
- Date of Issue:2022/05/23 13:12
- Category: inspeXio SMX-225CT
-
(SMX-225CT) System does not start even after turning the ...
Cause 1: Is the power supply circuitbreaker at the site powersupply service panel ormain power supply circuitbreaker on the back ofthe unit switched OFF? ... Show Detail
- No:8690
- Date of Issue:2022/05/23 13:19
- Category: inspeXio SMX-225CT
-
(SMX-225CT) The live image remains completely black
Cause 1: Are X-rays being emitted? Corrective Measures 1: Emit X-rays. Cause 2: Is the [OPERATE] keyswitch on the operationbox turned to [OFF]? Co... Show Detail
- No:8691
- Date of Issue:2022/05/23 13:30
- Category: inspeXio SMX-225CT
-
(SMX-225CT) Fluoroscopic images are blurry
Cause 1: Are the X-ray focussetting valuesappropriate? Corrective Measures 1:Adjust the focus values. Click here for "(SMX-225CT) Replacing the Filame... Show Detail
- No:8692
- Date of Issue:2022/05/23 13:35
- Category: inspeXio SMX-225CT
-
(SMX-225CT) A persistent image of the sample remains on t...
Cause: Has the same samplebeen scanned in the sameposition for a long time? Corrective Measures:Create an air calibration table. Click here for "(SMX-22... Show Detail
- No:8693
- Date of Issue:2022/05/23 13:38
- Category: inspeXio SMX-225CT
-
(SMX-225CT) X-ray generator trouble
«The interlockindicatorlamp isilluminatedred, in theX-ray controlarea in theCT softwareinspeXio64» Cause 1: Is the sliding door ormaintenance door on thes... Show Detail
- No:8694
- Date of Issue:2022/05/23 14:40
- Category: inspeXio SMX-225CT
-
(SMX-225CT) Problem with CT image
Cause 1:Does noise appear in theform of concentric circles? Corrective Measures 1:Click the [Calibration] buttonin the instrument settings areaof the... Show Detail
- No:8695
- Date of Issue:2022/05/23 14:48
- Category: inspeXio SMX-225CT
51-60 of Total 61