(PPSQ) Disassembling/Assembling the Reactor
1. Remove the inlet tube fixing screw. 2. Loosen the outlet tube fixing screw. NOTE: To Loosen a little is enough. Do not remove it. 3. Rem... Show Detail
- No:8418
- Date of Issue:2022/02/25 10:31
- Category: Protein Sequencer(PPSQ)
- Brached FAQ
-
When not using the system for a short time, it is sufficient to leave the pump unitpumping at a flow rate of 0.2 to 0.3mL/min and to turn OFF the power to th... Show Detail
- No:8419
- Date of Issue:2022/02/25 10:41
- Category: Protein Sequencer(PPSQ)
- Brached FAQ
-
(PPSQ) Pyridylethylation Sequence Schedule
The isocratic system allows pyridylethylation of cysteine in a PPSQ reactor precedingsequence analysis. The sequence schedule for pyridylethylation is insta... Show Detail
- No:8421
- Date of Issue:2022/02/25 11:30
- Category: Protein Sequencer(PPSQ)
-
(PPSQ) Samples with a High Salt Content
If a sample with a high salt content is analyzed with the protein sequencer, thebackground level in the chromatograms obtained may be high or the mobile phas... Show Detail
- No:8423
- Date of Issue:2022/02/25 12:29
- Category: Protein Sequencer(PPSQ)
-
(PPSQ) Analyzing Peptide Samples with PVDF Membranes
The sequence analysis of a peptide solution sample is usually performed using apolybrene-processed glass fiber disk as the sample support. With a PVDF membra... Show Detail
- No:8425
- Date of Issue:2022/02/25 12:35
- Category: Protein Sequencer(PPSQ)
-
O-rings are used to seal reagent/solvent bottles and the waste bottle. If the O-ring in a reagent/solvent bottle is cracked or cut, the pressure may decreas... Show Detail
- No:8427
- Date of Issue:2022/02/25 12:43
- Last Updated:2023/12/20 13:16
- Category: Protein Sequencer(PPSQ)
-
(PPSQ) Troubleshooting PTH-Amino Acid Analysis (Isocratic...
« Analysis takes longer to complete» • K is close to I and separation is poor. →The column temperature is too low.: Check the column temperature.(40ºC)... Show Detail
- No:8429
- Date of Issue:2022/02/25 13:22
- Category: Protein Sequencer(PPSQ)
-
Initial percent yield and average repetitive percent yield If the initial percent yield is "a" and the repetitive percent yield is "b", the percent yield "y... Show Detail
- No:8449
- Date of Issue:2022/03/01 13:13
- Category: Protein Sequencer(PPSQ)
-
Calculates the evaluation value as an indicator of peak fluctuations for each amino acid in chromatograms for consecutive cycles. The one with the highest ev... Show Detail
- No:8451
- Date of Issue:2022/03/01 13:21
- Category: Protein Sequencer(PPSQ)