(MCE) (ERROR)Current accuracy error
<Typical Causes> The HV voltage monitor value has fallen outside the specified range. <Instrument Operations> The microchip that generated... Show Detail
- No:8384
- Date of Issue:2022/02/18 10:23
- Last Updated:2022/02/25 15:51
- Category: MCE-202(MultiNA)
-
(PPSQ) In an Emergency / Actions in the Event of a Power ...
«In an Emergency / » If any problem is detected, such as a burning smell, take the following action. 1. Turn OFF the power of the PPSQmain unit. 2. Turn... Show Detail
- No:8395
- Date of Issue:2022/02/24 10:41
- Category: Protein Sequencer(PPSQ)
-
The procedure for replacing reagents is described below. After replacement, it isrecommended that the reagent and solvent bottles are purged in order to prev... Show Detail
- No:8398
- Date of Issue:2022/02/24 12:52
- Category: Protein Sequencer(PPSQ)
-
To replace a reagent without mixing the new reagent with the previous one, or whenshutting down the system, remove the reagent using the procedure described ... Show Detail
- No:8400
- Date of Issue:2022/02/24 13:09
- Last Updated:2022/02/24 15:58
- Category: Protein Sequencer(PPSQ)
-
(PPSQ) Replacing the mobile phase
1. Replacing the mobile phase 1)Click [LC Control] on the [Maintenance] menu in the [PPSQ Analysis] window. 2) Set the pump flow rate to 0mL/min ... Show Detail
- No:8402
- Date of Issue:2022/02/24 14:01
- Category: Protein Sequencer(PPSQ)
- Brached FAQ
-
(PPSQ) Replacing the Nitrogen Gas Cylinder
Occasionally check the nitrogen gas cylinder's pressure regulator during system operationand, if the residual pressure is low, replace the cylinder as soon a... Show Detail
- No:8405
- Date of Issue:2022/02/24 14:24
- Category: Protein Sequencer(PPSQ)
-
(PPSQ) Disposing of Waste Liquid
As sample analysis is repeated, the amount of liquid in the waste bottle increases. Occasionally check the amount of liquid in the waste bottle and dispose ... Show Detail
- No:8407
- Date of Issue:2022/02/24 14:32
- Category: Protein Sequencer(PPSQ)
-
« Removing the column» 1. When the pump is running, push [pump] onthe pump. 2. Open the cover of the column oven. 3. Remove the tube connecting to the i... Show Detail
- No:8409
- Date of Issue:2022/02/24 14:54
- Category: Protein Sequencer(PPSQ)
- Brached FAQ
-
1. Turning ON Power 1) Turn ON the power to the detector and the column oven. 2) Turn ON the power to the PPSQ. 3) Turn ON the power to the PC and the ... Show Detail
- No:8410
- Date of Issue:2022/02/24 15:29
- Category: Protein Sequencer(PPSQ)
- Brached FAQ
-
(PPSQ) Changing the Number of Analysis
It's possible to change the number of analyses during the execution of sample analysis. 1. Click [Change Number of Cycles] on the registration frame of th... Show Detail
- No:8415
- Date of Issue:2022/02/25 09:06
- Category: Protein Sequencer(PPSQ)