(MCE) The peak detection position fluctuates
•The temperature of the separation buffer was not sufficiently at room temperature. →Allow 30 minutes or more to elapse the separation buffer at room temp... Show Detail
- No:8363
- Date of Issue:2022/02/16 16:24
- Last Updated:2022/02/25 16:11
- Category: MCE-202(MultiNA)
-
«None of the microchipsdetect peaks for markers orsamples.» • Incorrect usage of microchips, reagent,rinse water, glass rinse water bottle, orwaste contai... Show Detail
- No:8365
- Date of Issue:2022/02/17 15:39
- Category: MCE-202(MultiNA)
-
(MCE) Difference in peak detection sensitivity
• The microchip surfaces arecontaminated. →If the surfaces are contaminated, rinse them. Click here for "(MCE) Inspecting and Washing Microchip Reserv... Show Detail
- No:8367
- Date of Issue:2022/02/17 15:51
- Category: MCE-202(MultiNA)
-
(MCE) There is a problem with peak recognition
«Peaks other than themarker peaks arerecognized as markerpeaks.» •Another peak near the marker ismistakenly recognized as the marker. →Set the correct ... Show Detail
- No:8369
- Date of Issue:2022/02/17 15:58
- Last Updated:2022/02/25 16:07
- Category: MCE-202(MultiNA)
-
(MCE) There is a problem with Size Accuracy
«Analysis performancedeviates from specificationaround the lower end of the size range.» •The ladder size and target fragmentsize are too far separated fr... Show Detail
- No:8371
- Date of Issue:2022/02/17 15:58
- Last Updated:2022/02/25 16:04
- Category: MCE-202(MultiNA)
-
(MCE) (ERROR)Cannot continue Sample holder is not present
<Typical Causes> • The sample holder is not properly installed. • The sensor that detects the sample holder is broken. <Corrective Measures... Show Detail
- No:8374
- Date of Issue:2022/02/18 10:04
- Last Updated:2022/02/25 16:04
- Category: MCE-202(MultiNA)
-
(MCE) (ERROR)Cannot continue Chip # is not present
<Typical Causes> • The microchips specified in the [Chip Management] window are notproperly installed. • The sample probe is damaged. • The sensor t... Show Detail
- No:8376
- Date of Issue:2022/02/18 10:09
- Last Updated:2022/02/25 16:02
- Category: MCE-202(MultiNA)
-
(MCE) (ERROR)Chip # has an error
The test current value for the chip that generated "fill check NG" was alsooutside the standard range after the separation buffer refill. <Instru... Show Detail
- No:8378
- Date of Issue:2022/02/18 10:16
- Last Updated:2022/02/25 16:01
- Category: MCE-202(MultiNA)
-
(MCE) (ERROR)Overcurrent error
An excessive current value was detected. <Typical Causes> • A short circuit was caused by leaking separation buffer or samplesolution on the mi... Show Detail
- No:8380
- Date of Issue:2022/02/18 10:23
- Category: MCE-202(MultiNA)
-
(MCE) (ERROR)Undercurrent error
An insufficient current value has been detected. <Typical Causes> • The microchip was clogged. • Air bubbles were mixed into the separation ch... Show Detail
- No:8382
- Date of Issue:2022/02/18 10:23
- Last Updated:2022/02/25 15:57
- Category: MCE-202(MultiNA)
21-30 of Total 49