(EZ-Test) An optional device is not functioning
•Confirm whether the presence of the optional device was registered correctly.Also, some optional devices require restarting the testing machine. Switch theE... Show Detail
- No:7688
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:38
- Category: EZ-Test Series
-
(EZ-Test) The test force display indicates "---"
• The load cell is not being recognized. Temporarily switch the power OFF andconfirm that the calibration cable is connected properly, then switch the powerO... Show Detail
- No:7687
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:33
- Category: EZ-Test Series
-
(EZ-Test) None of the buttons work
• No buttons function while theEZ-X unit is starting up. Wait until it finishesstartup. Show Detail
- No:7686
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:32
- Category: EZ-Test Series
-
(EZ-Test) Some buttons don't work
• The types of buttons that can be operated is restricted during tests or whileconnected to a computer. Try again after the test is finished or afterdisconne... Show Detail
- No:7685
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:32
- Category: EZ-Test Series
-
(EZ-Test) A servo error occurred during jog operations
• The motor may be overloaded. Confirm there are no obstructions caughtbetween the crosshead and table. Also confirm that nothing is wrapped aroundthe ball s... Show Detail
- No:7684
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:32
- Category: EZ-Test Series
-
(EZ-Test) A servo error occurred during testing
• The motormay be overloaded. Confirm whether the testing machine operatinglimits have been exceeded. (Max 10 hours continuous operation, max 1000cycles, or ... Show Detail
- No:7683
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:31
- Category: EZ-Test Series
-
• Confirm whether the emergency stop button is ON. Click here for "(EZ-Test) Safety Systems". • Confirm whether a stroke limit was triggered. The moto... Show Detail
- No:7682
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:31
- Category: EZ-Test Series
-
(EZ-Test) A ground fault circuit interrupter was tripped ...
• There may be an electrical leak within theEZ-X unit. Immediately stop usingthe system and contact a Shimadzu sales or service representative. Show Detail
- No:7681
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:30
- Category: EZ-Test Series
-
(EZ-Test) Fuses fail frequently
• Confirm whether the input voltage mode switch is set properly. • Confirm whether the fuses current rating is correct.(100V:10A, 220V:5A) • Confirm whethe... Show Detail
- No:7680
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:30
- Category: EZ-Test Series
-
(EZ-Test) Cannot switch the instrument power ON
• Confirm whether a circuit breaker has tripped on the power supply panel. • Confirm whether the power cordis disconnected. • Confirm whether the input vol... Show Detail