• Confirm whether the input voltage mode switch is set properly.
• Confirm whether the fuses current rating is correct. (100V/120V: 10 A, 220V/240V: 5A)
• Confirm whether a time delayed type of fuse is being used. (A "fast acting" fuse cannot be used.)