• No : 7707
  • Date of Issue : 2022/01/20 14:30
  • Print

(EZ-Test) Inspection of Stroke Speed

(EZ-Test) Inspection of Stroke Speed
Category : 

Answer

1. Display the jog speed setting screen by selecting the [System] - [Jog Speed], and make a note of the current jog speed.
2. Set the jog speed to 6mm/min.
3. Display the Testing screen by pressing the [Test] button.
4. Press the stroke zero reset button to set the stroke value to zero.
5. Perform the jog low speed operation for 10 seconds.
6. Check the stroke value. If the value is around 1.000mm or -1.000 mm, it is normal.
7. Set the jog speed back to the previous value in the jog speed setting screen by selecting the [System] - [Jog Speed].
If the stroke speed is abnormal, contact a Shimadzu sales or service representative.