Products Search
>
Materials Testing & Inspection
>
Autograph(AG)
>
EZ-Test Series
>
(EZ-Test) Fuses fail frequently
Back
No : 7680
Date of Issue : 2020/12/02 14:16
Last Updated : 2022/01/26 10:30
Print
(EZ-Test) Fuses fail frequently
(EZ-Test) Fuses fail frequently
Category :
Products Search
>
Materials Testing & Inspection
>
Autograph(AG)
>
EZ-Test Series
Answer
• Confirm whether the input voltage mode switch is set properly.
• Confirm whether the fuses current rating is correct. (100V: 10A, 220V: 5A)
• Confirm whether a time delayed type of fuse is being used. (A "fast acting" fuse cannot be used.)