(EHF) Step settings cannot be configured (program function test)

(EHF) Step settings cannot be configured (program function test)
During test method setup (program function test) Software Trouble
Windows Software for 4830
Category : 

Answer

• [Control change] is always set in the first step. This setting cannot be deleted or moved.
• An external input waveform can only be used for the final step.
• Multiple random waves cannot be set.
• Multiple waveform files cannot be set.
• Level trigger and digital trigger are only availabl e when a single controller is used.
• The cycle trigger and loop trigger cannot be combined.
• Multip le time triggers cannot be set.
• Multiple cycle triggers cannot be set up.
• Multiple loop triggers cannot be set up.
• Loops cannot be nested (one loop c annot be set up inside another loop).
• When the control TD is changed, a confirmation message appears and then all steps after the current step are deleted