(EHF) When automatic PID setting is performed, the piston leaves the test piece
(EHF) When automatic PID setting is performed, the piston leaves the test piece
During testing Software Trouble
Windows Software for 4830
Category :
Products Search
>
Materials Testing & Inspection
>
Servopulser
>
Windows Software for 4830
Answer
When obtaining the test force gain for a compression or bending test , preload the te st piece. During auto tuning, I-ACT is shut off, and if the compression force is small, the piston may leave the test piece.