The zero adjustment of the test force is executed automatically by a touch-panel operation (Auto Zero function).
CAUTION:
• The zero adjustment of the test force must be executed at the raised ram position where “READY” is shown normally (that is, the position where the ram is raised about 10 mm from its lowest position).
• Before executing the test-force cal1ibration, make sure that no load is applied.
• When Auto Zero is in progress, do not touch other keys.
1. To execute the Auto Zero function, touch the [→0←] key in the test-force display area of the LCD display.
NOTE: If you execute the Auto Zero with the upper grip and the lower grip closed, a value different from actual test force may be displayed.