Trouble1: Abnormality in measured light intensity data
Cause1: Sample is adhered to lens and/or filters
Counter measures1: Clean the measuring chamber
Trouble2: Air valve does not open
Cause2-1: Filter of dust collector is plugged
Counter measures2-1: Clean the dust collector filter
Cause2-2: Disconnection of cables and/or tubes
Counter measures2-2: Connect correctly
Cause2-3: Wrong insert position of the dust collector pipe
Counter measures2-3: Set the dust collector pipe inserting it to the upper one of the two receptacles
Cause2-4: Plugging of pressure switch filter and others
Counter measures2-4: Exchange the filter element and/or clean the silicone tube
Click here for "(SALD) Dry measurement case(SALD-DC23), Replace the Filter element".
Trouble3: Frequent fuse cuttings
Cause3: Sampling unit Presser regulator unit