• No : 7460
  • Date of Issue : 2021/11/18 13:27
  • Print

(SALD) High-Concentration Measurement Unit, Troubleshooting

(SALD) High-Concentration Measurement Unit, Troubleshooting
SALD-HC23S / SAL-HC75S
Category : 

Answer

Problem: Photometric data shows abnormalities
 
Cause1: セル内に気泡が付着している。
Corrective Action1: Make sample without bubbles.
  Click here for "(SALD) High-Concentration Measurement Unit, Sample Creation Procedure".
 
Cause2: Soiled slide glass
Corrective Action2: Wipe the slide glass softly by the lint-free wiping cloth for industrial use.
  Click here for "(SALD) High-Concentration Measurement Unit, Cleaning the cell/cell holde".
 
Cause3: Defect install of the slide glass
Corrective Action3: Insert the slide glass to the cell holder correctly.
  Click here for "(SALD) High-Concentration Measurement Unit, Install/Operation".
 
Cause4: Defect install of the cell holder
Corrective Action4: Install the cell holder and lock it correctly.
  Click here for "(SALD) Removing/Installing the Cell Holder".