Problem1: Photometric data shows abnormalities
Cause1-1: Bubbles adhering to the cell
Corrective Action1-1: Clean the cell using the same type of solvent. Should bubbles persist, clean the flow cell.
Click here for "(SALD) Sampler, Cleaning of the flow cell".
Cause1-2: Soiled cell
Corrective Action1-2: When testing samples that adhere easily, the cell may get soiled during measurement causing prejudice in data. Clean the flow cell.
Click here for "(SALD) Sampler, Cleaning of the flow cell".
Problem2: Liquid leaks
Cause2-1: Faulty cell assembly
Corrective Action2-1: The flow cell may not be sealed completely. Reassemble a flow cell.
Click here for "(SALD) Sampler, Cleaning of the flow cell".
Cause2-2: Cell chipped
Cause2-3: Faulty sample bath assembly
Corrective Action2-3: Reassemble sample bath.
Click here for "(SALD) Sampler, Cleaning of the sample bath".