Remove the cell from the cell holder before performing removal or attachment of the cell holder.
Refer to the instruction manual of each option unit for how to remove the cell.
CAUTION: If removing or installing the cell holder, turn OFF the power of instrument.
Not doing so may result in ignition, electric shock, or injury.
1. Removing the Cell Holder
1) Pull out the plug inserted in the cell holder and insert the plug in the connector at the left upper part in a measuring chamber.
2) Push the removal button in front side of the measuring chamber, and raise a little the front side of a cell holder.
3) Take out a cell holder from a measuring chamber.
NOTE: When you take out a cell holder, be careful to damage neither a filter nor a lens.
2. Installing the Cell Holder
1) Insert the locating hole of the cell holder rear to the locating pin in a measuring chamber.
NOTE: When you install a cell holder, be careful to damage neither a filter nor a lens.
2) Take down the front of a cell holder slowly, and insert a cell holder in a lock mechanism and fix.