(GAS) The indicator drifts (CGT-7100)
Back
No : 6822
Date of Issue : 2021/08/20 11:25
Print
(GAS) The indicator drifts (CGT-7100)
(GAS) The indicator drifts (CGT-7100)
Category :
Products Search
>
Continuous Monitoring Analysis
Answer
• There is a large change in ambient temperature. → Select a location with as little possible change in temperature.
• The inner surface of the cell is dirty. → Clean and purge the measurement cell. Contact
a Shimadzu sales or service representative
.
• Problem in sampling unit. → Perform checks and repairs on each part.
• Gas is leaking inside the analyzer. → Check the leak in the piping.
Click here
for "Gas leak checking (CGT-7100)".