Products Search
>
GC
>
GC-2030
>
Error/Warning
>
AOC-20s vial returning error[Error Code:6019]
Back
No : 6063
Date of Issue : 2021/05/18 15:33
Last Updated : 2021/06/16 09:22
Print
AOC-20s vial returning error[Error Code:6019]
The following causes are possible.
• Gripper could not grip the vial when returning it and dropped the vial at the handover position.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Check and perform the following.
• Redo the teaching using the vial to be used for actual analysis.
[In case of AOC-30i]
AOC-30i_AOC-20sU_SamplerTeaching.pdf
[In case of AOC-20i]
AOC-20i_AOC-20s_SamplerTeaching.pdf
• Perform a transport test (in case of AOC-30i).
AOC-30i_AOC-20sU_TransportTest.pdf