AOC-20s vial setting error[Error Code:6018]
The following causes are possible.
• A vial in the handover position was not removed before starting analysis.
• Grip was not firm enough. Vial dropped when it was being lifted.
• Vial hit an obstacle during transport and ended up dropping.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Check and perform the following.
• Check for any vial left from the previous sampling.
• Redo the teaching using the vial to be used for actual analysis.
[In case of AOC-30i]
AOC-30i_AOC-20sU_SamplerTeaching.pdf
[In case of AOC-20i]
AOC-20i_AOC-20s_SamplerTeaching.pdf
• Perform a transport test (in case of AOC-30i).
AOC-30i_AOC-20sU_TransportTest.pdf