The following causes are possible.
• The lifting motor has a problem.
• The deviation of the installation position from the teaching result is causing the gripper to hit the vial.
• There is an obstacle in the path to the destination, which is interfering with the vial.
• A vial with dimensions different from those of the vial used for teaching is used.