Products Search
>
GC
>
GC-2030
>
Error/Warning
>
AOCm installation error[Error Code:6010 6032]
Back
No : 6051
Date of Issue : 2021/05/18 15:33
Last Updated : 2021/06/11 14:43
Print
AOCm installation error[Error Code:6010 6032]
The following causes are possible.
• The mounting guide in the back and to the far left is not in contact with the installation sensor.
• The installation sensor has a problem.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Check that the auto injector is mounted correctly and not at an angle.